Emerson HASAxE-IM-HS Welding Consumables User Manual


 
X-STREAM
Emerson Process Management GmbH & Co. OHG 6-7
Instruction Manual
HASAxE-IM-HS
05/2006
6 Initial Startup
6-2 Checking the Instrument's Setup
Analog output..
PumpControl: Menu
Note!
If an internal pump is installed, the menu
shows two lines, without pump the lower line
does not show.
The line "PumpControl" allows to specify how
the optional internal pump is controlled
(switched ON/OFF): Either by a software Menu
or by an external signal, applied via Digital
Input 7, if such is installed (see INSTALLED
OPTIONS menu).
CrossInterference
Status: Off
@ Span Cal: No
The first line allows to enable (On) or disable
(Off) two-way cross interference compen-
sation of two channels. Enabled compensation
minimizes the interacting influences of the
measuring gases in both channels.
The second line specifies if the cross com-
pensation factors are automatically calculated
during a next span calibration or not. "Yes"
enables automatic calculation but requires
using pure gases for one single span cali-
bration procedure.
For a detailed description of how to setup cross
compensation 7-3-7, page 7-22.
Press the LEFT key to return to the SETUP
screen and enter the IN/OUTPUTS menu.
6-2-4 Analog Output Setup