Emerson HASAxE-IM-HS Welding Consumables User Manual


 
X-STREAM
Emerson Process Management GmbH & Co. OHG7-10
Instruction Manual
HASAxE-IM-HS
05/2006
7-3-3 Advanced Calibration
7-3-3 Advanced Calibration
Standard manual calibration procedures offer
limited funtionality:
To zero and span calibrate a dual channel
instrument the operator has to manually start 4
procedures in proper sequence. In addition he
has to stay at the instrument to see when the
one sequence has finished and to start the
following.
The same is applicable for a single channel
instrument, when the operator wants to perform
both zero and span calibrations.
To improve even manual calibration
procedures, X-STREAM analyzers offer a new
ADVANCED CALIBRATION menu: It allows
single key activation for
zero calibration of both channels of an
analyzer
span calibration of both channels of an
analyzer
zero and span calibration of both
channels of an analyzer
Although advanced calibration offers most
advantage for dual channel instruments, is may
be used for single channel analyzers as well:
Activate zero and span calibration for the one
channel by a single key.
The only precondition for making use of this
new feature is to have installed internal and/or
external valves and proper assigned.
For a description of how to
prepare advanced calibrations
7-3-3-1, page 7-11
perform advanced zero calibrations 7-3-3-2, page 7-14
perform advanced span calibrations 7-3-3-3, page 7-16
perform advanced zero & span calibrations 7-3-3-4, page 7-18