Emerson HASAxE-IM-HS Welding Consumables User Manual


 
X-STREAM
Emerson Process Management GmbH & Co. OHG9-14
Instruction Manual
HASAxE-IM-HS
05/2006
9-3 Modbus - List of Parameters and Registers
Attribute
1
Daniel /
Enron
Modicon Type
Range / Enum values / Coefficients
Modbus-Register
Mnemonic
Description
3139
3140
3139
3140
TempCheckEnable#DB
defines if a channel is checked for correct
temperature range
0 = Off
1 = use Temperature 1
2 = use Temperature 2
Cfg, Fac
Enum
3012
3014
3012
3014
TempCompSpan#DB
activates temperature span compensation
procedure for concentration measurement
0 = Off
1 = use Temperature 1
2 = use Temperature 2
Cfg, Fac
Enum
3011
3013
3011
3013
TempCompZero#DB
activates temperature offset compensation
procedure for concentration measurement
0 = Off
1 = use Temperature 1
2 = use Temperature 2
Cfg, Fac
Enum
3080 3080
TempDecimalPointDB
decimal point position for temperature displays
0..4
Cfg
uchar
9027
9028
6053..6054
6055..6056
Temperature#DB
calculated temperature value in °C
Dyn, Sim
Float
9013
9014
6025..6026
6027..6028
TemperatureFactor#DB
span correction of temperature compensation
0.05-1.8, depending on temperature
Dyn, Sim
Float
9007
9008
6013..6014
6015..6016
TemperatureOffset#DB
zero correction of temperature compensation
Dyn, Sim
Float
3143
3144
3143
3144
TempHighLimit#DB
high limit level of the correct temperature range
value to be °C;
0..250
Cfg, Fac
Byte
3141
3142
3141
3142
TempLowLimit#DB
low limit level of the correct temperature range
value to be °C;
0..250
Cfg, Fac
Byte
1025
1026
1025
1026
TempRangeError#DB
temperature is outside the configured allowed
range
Dyn
Boolean
3079 3079
TempUnitDB
specify the temperature unit to show on front
panel display
0 = °C 1 = °F
Cfg
Enum
9055..9058
9059..9062
6109..6116
6117..6124
TfactCorrCoeffs#DB
polynom coefficients that compute span factor
correction (TemperatureFactor#DB) using the
assigned temperature (TempCompSpan#DB)
A0, A1, A2, A3
Cfg, Fac
Float-Arr4
9047..9050
9051..9054
6093..6100
6101..6108
ToffCorrCoeffs#DB
polynom coefficients that compute zero offset
correction (TemperatureOffset#DB) using the
assigned temperature (TempCompZero#DB)
A0, A1, A2, A3
Cfg, Fac
Float-Arr4
This list is ordered alphabetically by the parameter name („Mnemonic“) column.
Any "#" is a wildcard, to be replaced by "1" or "2" for the desired measuring channel. In this case
the upper register line is assigned to channel 1, the lower line to channel 2.