Emerson HASAxE-IM-HS Welding Consumables User Manual


 
X-STREAM
Instruction Manual
HASAxE-IM-HS
05/2006
1-2 Emerson Process Management GmbH & Co. OHG
1-1 Overview
The X-STREAM series analyzers offer a wide
range of available configurations and options,
to be combined according to the selected
model:
Measuring principles
Up to two out of all offered principles may be
combined within one analyzer model to provide
best adaption to the application.
For a detailled description of available
measuring principles: chapter 3.
Gas path design
Internal tubing with viton or, optional and
depending on application, PFA or stainless
steel.
In addition one or more of the following options
are available:
Soleniod valve block
This option uses 4 internal solenoid valves
to control sample, zero, span gas 1 and
span gas 2. These gases are fed to the
analyzers to provide manual or controlled
automatic calibration (initialized by
keypad, serial interface or digital inputs).
Sample pump
Maximum flow rate 2.5 l/min
Barometric Pressure sensor
(Measuring range 800 to 1,200 hPa)
Facilitates compensation of atmospheric
pressure variations to improve precision
of results (
measurement specifi-
cations, page 3-17).
Special sensors for e.g. corrosive gas on
request.
Flow measurement
A flow sensor (option) can be used to
monitor gas flow and set alarms.
Heated box for physical components
All the physical components
*)
can
optionally be installed inside thermostatted
box to minimize influences from ambient
tem-perature fluctuations.
For a detailled description of optional gas path
components: page 1-16
Interfaces
All models may be configured to use several
interfaces:
Standard:
analog outputs
serial interface (RS 485 or RS 232) with
Modbus protocol
status signals (NAMUR; relay outputs)
Optional:
8 digital outputs & 7 digital inputs
For a detailled description of optional inter-
faces:
page 1-18