6-2 Checking the Instrument's Setup
Signal
Range
Setting
Operation Mode
Failure Signal
Level acc. NE 43
Valid Measuring
Signal Range
Measuring Signal
Exceeds Lower
Range Limit
Measuring Signal
Exceeds Upper
Range Limit
Measuring
System
Failure
Cable
Break
Channel 2
when not
equipped
0-20mA
Dead-Zero - 0 … 20 mA < -19 mA > 21 mA undefined 0 mA 0 mA
4-20mA
Life-Zero - 4 … 20 mA < -19 mA > 21 mA undefined 0 mA 4 mA
LL0-20mA
similiar
Dead-Zero
below 0 … 20 mA -0,2 mA 20,5 mA -2 mA 0 mA -0.2 mA
LL4-20mA
similiar
Life-Zero
below 4 … 20 mA 3,8 mA 20,5 mA 2 mA 0 mA 3.8 mA
HL0-20mA
similiar
Dead-Zero
above 0 … 20 mA -0,2 mA 20,5 mA >21 mA 0 mA -0.2 mA
HL4-20mA
similiar
Life-Zero
above 4 … 20 mA 3,8 mA 20,5 mA >21 mA 0 mA 3.8 mA
Output Signal, when…
Table 6-1: Analog Output Signals Settings & Operation Modes
Dual channel instrument:
Pressing the key while any line is selected
returns to the optional gas component selec-
tion menu to open the same menu for the
other measuring channel.
Press the LEFT key to return to the ANALOG
SIGNAL SETUP screen, enter the "Scaling.."
line and select the channel to setup (multi-
channel instruments only).
Enter the gas concentrations generating ana-
log output signals according the lower
respectively upper range (4(0) or 20 mA). By
default the gas concentrations comply with the
measuring ranges. Setting other
concentrations allows to limit the output signal
range to a concentration level range smaller
than full scale.
Scaling
0/4mA: 0.0000 ppm
20mA: 2000.0 ppm