7 Maintenance
• Power off the analyzer and close the
housing.
In a next step now perform a zero and span
calibration at least for the channel the replaced
sensor is related to.
Close the worn-out sensor´s opening with the
sealing plug taken from the new sensor. Send it
back to the EMERSON Process Management
factory (or to your local sales office) or to an
industrial waste management contractor for
waste disposal.
7-4 Replacing the Electrochemical Sensor
7-4-6 Finalizing the Sensor Replacement