X-STREAM
5-29
Instruction Manual
HASAxE-IM-HS
05/2006
Emerson Process Management GmbH & Co. OHG
5 User Interface
5-4 Menu System - Measurement Setup Menu
5-4-3-3 Measurement Setup Menu
If system is setup accordingly access level 2
code must be entered to access this menu.
If no pressure sensor is installed (INSTALLED
OPTIONS menu - "PressSensor" set to
manual) enter the current ambient pressure
here.
Input range: 500 .. 2000 hPa
Otherwise this field is not editable and shows
the pressure sensor value.
Signal damping setup menu
5-4-3-3-1, page 5-30
Cross interference setup menu
5-4-3-3-2, page 5-31
Note 1!
The unit for the pressure value is taken from
the related entry in the display setup menu.
Note 2!
As the pressure value is used for pressure
compensation update it on a regular basis
when set to "manual" to achieve best measu-
ring results.
AD conversion setup
5-4-3-3-3, page 5-32
Setup..
Measurement..
Pressure: 1014.0 hPa
Damping..
CrossInterference..
AD conversion..