Emerson HASAxE-IM-HS Welding Consumables User Manual


 
X-STREAM
Emerson Process Management GmbH & Co. OHG6-12
Instruction Manual
HASAxE-IM-HS
05/2006
6-2 Checking the Instrument's Setup
Protocol: MODB RTU
MODB mode: 32Bit
ID number: 2
Interface: RS485/2w
Baud rate: 19200
Parity: None
Level1: 100 ppm
Function: Low
Level2: 500 ppm
Function: High
Press the LEFT key several times to return to
the SETUP MENU, select the "Alarms.." line,
enter the menu and select the channel to be
setup (multi-channel instruments only).
6-2-5 Concentration Alarms Setup
Two different concentration thresholds can be
set for each channel, freely configurable within
the limits given by the "SpanRange" parameter.
The moment the measured concentration
exceeds one of the thresholds, a status message
appears in the measuring screen's 4
th
line and
the related digital output (option) is set.
Press the LEFT key several times to return to the
SETUP MENU, select the "Communication.." line
and enter the menu.
Note!
If you do not want to make use of concentration
alarms, skip the next pages and continue with
page 6-18.