Agilent Technologies E1401-90493 Work Light User Manual


 
Chapter 2 209
Performance Verification Tests
24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B
continue with step 15.
9. Note the most positive number from the Flatness Relative to 50 MHz
(dc) column of Table 2-50 for frequencies between 800 MHz and
1.0 GHz. Record this number in Table 2-55 as the Maximum
Response for Band 0A.
10.Note the most negative number from the Flatness Relative to
50 MHz (dc) column of Table 2-50 for frequencies between 800 MHz
and 1.0 GHz. Record this number in Table 2-55 as the Minimum
Response for Band 0A.
11.For Band 0A in Table 2-55, subtract the Minimum Response value
from the Maximum Response value and record the result in the
Peak-to-Peak Response column.
12.Note the most positive number from the Flatness Relative to 50 MHz
(dc) column of Table 2-50 for frequencies between 1.7 GHz and
2.0 GHz. Record this number in Table 2-55 as the Maximum
Response for Band 0B.
13.Note the most negative number from the Flatness Relative to
50 MHz (dc) column of Table 2-50 for frequencies between 1.7 GHz
and 2.0 GHz. Record this number in Table 2-55 as the Minimum
Response for Band 0B.
14.For Band 0B in Table 2-55, subtract the Minimum Response value
from the Maximum Response value and record the result in the
Peak-to-Peak Response column.
15.Note the most positive number from the Flatness Relative to 50 MHz
(dc) column of Table 2-52 for frequencies between 3.0 GHz and
6.7 GHz. Record this number in Table 2-55 as the Maximum
Response for Band 1.
16.Note the most negative number from the Flatness Relative to
50 MHz (dc) column of Table 2-52 for frequencies between 3.0 GHz
and 6.7 GHz. Record this number in Table 2-55 as the Minimum
Response for Band 1.
17.For Band 1 in Table 2-55, subtract the Minimum Response value
from the Maximum Response value and record the result in the
Peak-to-Peak Response column.
18.Note the most positive number from the Flatness Relative to 50 MHz
(dc) column of Table 2-52 for frequencies between 6.7 GHz and
13.2 GHz. Record this number in Table 2-55 as the Maximum
Response for Band 2.
19.Note the most negative number from the Flatness Relative to
50 MHz (dc) column of Table 2-52 for frequencies between 6.7 GHz
and 13.2 GHz. Record this number in Table 2-55 as the Minimum
Response for Band 2.