Agilent Technologies E1401-90493 Work Light User Manual


 
252 Chapter 2
Performance Verification Tests
29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B,
E4407B, and E4408B
frequencies listed in Table 2-68 for Band 1.
NOTE End of procedure for Agilent E4404B.
Band 2
17.On the analyzer, press the following keys:
FREQUENCY, Center Freq, 9 GHz
18.On the synthesized sweeper, press CW, 9 GHz.
19.Enter the power sensor 9 GHz calibration factor into the power
meter.
20.On the analyzer press the following keys:
AMPLITUDE, Presel Center
Marker
, More, Marker All Off
21.Repeat step 5 through step 11 for the synthesized sweeper CW
frequencies listed in Table 2-68 for Band 2.
NOTE End of procedure for Agilent E4405B.
Band 3
22.On the analyzer, press the following keys:
FREQUENCY, Center Freq, 15 GHz
23.On the synthesized sweeper, press: CW, 15 GHz.
24.Enter the power sensor 15 GHz calibration factor into the power
meter.
25.On the analyzer press the following keys:
AMPLITUDE, Presel Center
Marker
, More, Marker All Off
26.Repeat step 5 through step 11 for the synthesized sweeper CW
frequencies listed in Table 2-68 for Band 3 for the 15 GHz analyzer
center frequency.
Band 4
27.On the analyzer, press:
FREQUENCY, Center Freq, 21 GHz
28.On the synthesized sweeper, press: CW, 21 GHz.
29.Enter the power sensor 21 GHz calibration factor into the power
meter.