Agilent Technologies E1401-90493 Work Light User Manual


 
Chapter 2 253
Performance Verification Tests
29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B,
E4407B, and E4408B
30.On the analyzer, press AMPLITUDE, Presel Center.
31.On the analyzer, press the following:
Marker, More, Marker All Off
32.Repeat step 5 through step 11 for the synthesized sweeper CW
frequencies listed in Table 2-68 for Band 4 for the 21 GHz analyzer
center frequency.
Table 2-68 Other Input-Related Spurious Responses
Band
Analyzer
Center
Frequency
(GHz)
Synthesized
Sweeper CW
Frequency
(MHz)
Mixer
Level
(dBm)
Test Record
Entry
Mkr1 or
Mkr2
Amplitude
02.0
2042.8
a
–20 1)
2.0
2642.8
a
–20 2)
2.0
1820.8
b
–20 3)
2.0
278.5
b
–20 4)
Note: The following data applies only to the Agilent E4404B, E4405B,
E4407B and E4408B
02.0
5600.0
c
–10 5)
2.0
6242.8
c
–10 6)
14.0
4042.8
a
–20 7)
4.0
4642.8
a
–20 8)
4.0
3742.9
b
–20 9)
4.0
2242.8
c
–10 10)
Note: The following data applies only to the Agilent E4405B, E4407B and
E4408B
29.0
9042.8
a
–20 11)
9.0
9642.8
a
–20 12)
9.0
4982.1
b
–20 13)
9.0
9342.8
c
–10 14)
Note: The following data applies only to the Agilent E4407B and E4408B
315.0
15042.8
a
–20 15)