Agilent Technologies E1401-90493 Work Light User Manual


 
Chapter 2 245
Performance Verification Tests
28. Other Input-Related Spurious Responses: Agilent E4401B and E4411B
75 Input only: Use the minimum loss pad and 75 adapters to
connect to the 75
power sensor.
4. Adjust the power level of the synthesized sweeper for a
10 dBm
±0.1 dB reading on the power meter.
5. On the synthesized sweeper, press
SAVE, 1.
Figure 2-42 Other Input Related Spurious Responses Power Setting Setup
6. Set the CW frequency on the synthesized sweeper to 510.7 MHz.
7. Adjust the synthesized sweeper power level for a
10 dBm ±0.1 dB
reading on the power meter.
8. On the synthesized sweeper, press
SAVE 2.
9. Enter the 1 GHz calibration factor for the power sensor into the
power meter.
10.Set the CW frequency on the synthesized sweeper to 1310.7 MHz.
11.Adjust the synthesized sweeper power level for a
10 dBm ±0.1 dB
reading on the power meter.
12.On the synthesized sweeper, press
SAVE 3.
13.Enter the 100 MHz calibration factor of the power sensor into the
power meter.
14.Set the CW frequency of the synthesized sweeper to 100 MHz.
15.Adjust the synthesized sweeper power level for a
10 dBm ±0.1 dB
reading on the power meter.