Performance Verification Test Records
Agilent E4404B Performance Verification Test Record
Log, Res BW 300 Hz
(Option 1DR only)
Reference Level
–10 dBm –0.3 dB (17)________ 0.3 dB
±0.091 dB
0 dBm –0.3 dB (18)________ 0.3 dB
±0.100 dB
–30 dBm –0.3 dB (19)________ 0.3 dB ±0.092 dB
–40 dBm –0.3 dB (20)________ 0.3 dB
±0.094 dB
–50 dBm –0.5 dB (21)________ 0.5 dB
±0.095 dB
–60 dBm –0.5 dB (22)________ 0.5 dB
±0.097 dB
–70 dBm –0.5 dB (23)________ 0.5 dB
±0.099 dB
–80 dBm –0.7 dB (24)________ 0.7 dB
±0.101 dB
Linear, Res BW
300 Hz
(Option 1DR only)
Reference Level
–10 dBm –0.3 dB (25)________ 0.3 dB
±0.091 dB
0 dBm –0.3 dB (26)________ 0.3 dB
±0.100 dB
–30 dBm –0.3 dB (27)________ 0.3 dB ±0.092 dB
–40 dBm –0.3 dB (28)________ 0.3 dB
±0.094 dB
–50 dBm –0.5 dB (29)________ 0.5 dB
±0.095 dB
–60 dBm –0.5 dB (30)________ 0.5 dB ±0.097 dB
–70 dBm –0.5 dB (31)________ 0.5 dB
±0.099 dB
–80 dBm –0.7 dB (32)________ 0.7 dB
±0.101 dB
16. Resolution Bandwidth
Switching Uncertainty
Resolution Bandwidth
3 kHz –0.3 dB (1)________ 0.3 dB
±0.048 dB
9 kHz –0.3 dB (2)________ 0.3 dB ±0.048 dB
Table 3-8 Agilent E4404B Performance Verification Test Record
Agilent Technologies
Model E4404B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty