Chapter 3 471
Performance Verification Test Records
Agilent E4402B Performance Verification Test Record
33. Gain Compression
Test Frequency
53 MHz (1)________ 1.0 dB
±0.123 dB
50.004 MHz
(Option 1DR only)
(2)________ 1.0 dB ±0.123 dB
1403 MHz (3)________ 1.0 dB
±0.123 dB
2503 MHz (4)________ 1.0 dB
±0.123 dB
36. Displayed Average
Noise Level
Note: Enter results with preamp on in the appropriate section based
upon the ambient temperature when the test was performed.
1 kHz RBW, Preamp Off
10 MHz to 1 GHz (1)________ –117 dBm
±1.23 dB
1 GHz to 2 GHz (2)________ –116 dBm
±1.23 dB
2 GHz to 3 GHz (3)________ –114 dBm
±1.23 dB
1kHz RBW, Preamp On,
0 to 55
° C
10 MHz to 1 GHz (4)________ –132 dBm
±1.23 dB
1 GHz to 2 GHz (5)________ –132 dBm
±1.23 dB
2 GHz to 3 GHz (6)________ –129 dBm
±1.23 dB
10 Hz RBW, Preamp Off
10 MHz to 1 GHz (7)________ –136 dBm
±1.23 dB
1 GHz to 2 GHz (8)________ –135 dBm
±1.23 dB
2 GHz to 3 GHz (9)________ –133 dBm ±1.23 dB
10 Hz RBW, Preamp On,
0 to 55
° C
10 MHz to 1 GHz (10)________ –151 dBm
±1.23 dB
1 GHz to 2 GHz (11)________ –151 dBm ±1.23 dB
2 GHz to 3 GHz (12)________ –148 dBm
±1.23 dB
Table 3-4 Agilent E4402B Performance Verification Test Record
Agilent Technologies
Model E4402B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty