Chapter 2 347
Performance Verification Tests
43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)
Tracking Generator Level Flatness, Center
Frequency ≤100 kHz
NOTE Perform step 1 to step 7 for 50 Ω tracking generators only
(Option 1DN).
1. Set up the digital multimeter as follows.
NOTE Perform step 2 to step 7 for each measurement value in Table 2-97.
2. Refer to Figure 2-62 to set up the equipment.
Figure 2-62 Tracking Generator Level Flatness Test Setup,
≤100 kHz
3. To set the analyzer center frequency to 9 kHz, press
FREQUENCY,
9kHz (or as indicated in Table 2-96).
Parameter Setting
AC/DC AC Volts
Impedance & Units:
Set to 50
Ω impedance
SMATH 10
a
Set to dBm
MATH 5
a
Set to Synchronous
SETACV 3
a
Sub-sampled mode
a. To set the Agilent 3458A multimeter functions from the
front panel, press the blue shift key, then Recall State (T)
key. Use the
⇑ (up) and ⇓ (down) arrows to select the
appropriate function, then enter the value from the
numeric keypad and press enter.