514 Chapter 3
Performance Verification Test Records
Agilent E4404B Performance Verification Test Record
Peak-to-Peak Response (21)________ 3.0 dB ±0.39 dB
27. Frequency Response
(Preamp On)
(Option 1DS Only)
Note: Enter data in the appropriate section below depending upon
the ambient temperature at which the test was performed and
installed options.
20 to 30
° C
Band 0
(1 MHz to 3.0 GHz)
Maximum Response (1)________ 1.5 dB
±0.28 dB
Minimum Response –1.5 dB (2)________
±0.28 dB
Peak-to-Peak Response (3)________ 3.0 dB
±0.28 dB
Band 0A
(800 MHz to 1.0 GHz)
(serial number
US39440498 or greater)
Maximum Response (4)________ 1.5 dB
±0.28 dB
Minimum Response –1.5 dB (5)________
±0.28 dB
Peak-to-Peak Response (6)________ 3.0 dB
±0.28 dB
Band 0B
(1.7 GHz to 2.0 GHz)
(serial number
US39440498 or greater)
Maximum Response (7)________ 1.5 dB
±0.28 dB
Minimum Response –1.5 dB (8)________
±0.28 dB
Peak-to-Peak Response (9)________ 3.0 dB ±0.28 dB
Table 3-8 Agilent E4404B Performance Verification Test Record
Agilent Technologies
Model E4404B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty