Agilent Technologies E1401-90493 Work Light User Manual


 
Chapter 2 363
Performance Verification Tests
46. Tracking Generator Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B,
E4405B, E4407B, and E4408B (Option 1DN)
f. Press MKR, MARKERS (Off).
8. From Table 2-100, enter the 2nd Harmonic Level at 9 kHz as Test
Record entry 1 and copy this value into the performance verification
test record.
Test Record entry 1:
TG 2nd Harmonic Spurious Output __________ dB
9. From Table 2-100, locate the most positive 2nd Harmonic Level for
tracking generator frequencies of 9 kHz to 750 MHz and record this
value as Test Record entry 2 and copy this value into the
performance verification test record.
Test Record entry 2:
TG 2nd Harmonic Spurious Output __________ dB
10.From Table 2-100, enter the 3rd Harmonic Level at 9 kHz as Test
Record entry 3 and copy this value into the performance verification
test record.
Test Record entry 3:
TG 3rd Harmonic Spurious Output __________ dB
11.From Table 2-100, locate the most positive 3rd Harmonic Level for
tracking generator frequencies of 25 kHz to 1500 MHz and record
this value as Test Record entry 4 and copy this value into the
performance verification test record.
Test Record entry 4:
TG 3rd Harmonic Spurious Output __________ dB
12.Press
System, Alignments, Auto Align, All.
Table 2-100 Tracking Generator Harmonic Spurious Response Worksheet
1.5 GHz Tracking
Generator Output
Frequency
2nd Harmonic Level
(dBc)
3rd Harmonic Level
(dBc)
9 kHz
25 kHz
100 MHz
300 MHz
900 MHz
1500 MHz N/A