Agilent Technologies E1401-90493 Work Light User Manual


 
594 Chapter 3
Performance Verification Test Records
Agilent E4407B Performance Verification Test Record
GSM Transmit Power
Accuracy
(Option BAH only)
GSM Band
Input Amplitude
15 dBm –0.94 dB (13)________ 0.94 dB
±0.173 dB
20 dBm –0.94 dB (14)________ 0.94 dB ±0.161 dB
30 dBm –0.80 dB (15)________ 0.80 dB ±0.156 dB
40 dBm –0.83 dB (16)________ 0.83 dB ±0.161 dB
50 dBm –0.99 dB (17)________ 0.99 dB ±0.152 dB
60 dBm –1.13 dB (18)________ 1.13 dB ±0.161 dB
DCS and PCS Bands
Input Amplitude
15 dBm –0.92 dB (19)________ 0.92 dB
±0.173 dB
20 dBm –0.92 dB (20)________ 0.92 dB ±0.161 dB
30 dBm –0.74 dB (21)________ 0.74 dB ±0.156 dB
40 dBm –0.79 dB (22)________ 0.79 dB ±0.161 dB
50 dBm –0.95 dB (23)________ 0.95 dB ±0.152 dB
−60 dBm –1.09 dB (24)________ 1.09 dB ±0.161 dB
cdmaOne Receive
Channel Power, Preamp
Off
(Option BAC only)
Cellular Band
Input Amplitude
40 dBm –1.56 dB (25)________ 1.56 dB ±0.161 dB
60 dBm –1.56 dB (26)________ 1.56 dB ±0.161 dB
70 dBm –1.56 dB (27)________ 1.56 dB ±0.161 dB
80 dBm –1.56 dB (28)________ 1.56 dB ±0.161 dB
Table 3-12 Agilent E4407B Performance Verification Test Record
Agilent Technologies
Model E4407B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty