Chapter 3 549
Performance Verification Test Records
Agilent E4405B Performance Verification Test Record
1403 MHz (3)________ 1.0 dB ±0.123 dB
2503 MHz (4)________ 1.0 dB
±0.123 dB
4403 MHz (5)________ 1.0 dB
±0.123 dB
7603 MHz (6)________ 1.0 dB ±0.123 dB
37. Displayed Average
Noise Level
Note: Enter results with preamp on in the appropriate section based
upon the ambient temperature when the test was performed.
1 kHz RBW, Preamp Off
10 MHz to 1 GHz (1)________ –116 dBm
±1.23 dB
1 GHz to 2 GHz (2)________ –116 dBm
±1.23 dB
2 GHz to 3 GHz (3)________ –112 dBm
±1.23 dB
3 GHz to 6 GHz (4)________ –112 dBm
±1.23 dB
6 GHz to 12 GHz (5)________ –111 dBm
±1.23 dB
12 GHz to 13.2 GHz (6)________ –107 dBm
±1.23 dB
1 kHz RBW, Preamp On,
0 to 55
° C
10 MHz to 1 GHz (7)________ –131 dBm
±1.23 dB
1 GHz to 2 GHz (8)________ –131 dBm
±1.23 dB
2 GHz to 3 GHz (9)________ –127 dBm ±1.23 dB
10 Hz RBW, Preamp Off
10 MHz to 1 GHz (10)________ –135 dBm
±1.23 dB
1 GHz to 2 GHz (11)________ –135 dBm ±1.23 dB
2 GHz to 3 GHz (12)________ –131 dBm
±1.23 dB
3 GHz to 6 GHz (13)________ –131 dBm
±1.23 dB
6 GHz to 12 GHz (14)________ –130 dBm ±1.23 dB
12 GHz to 13.2 GHz (15)________ –120 dBm
±1.23 dB
Table 3-10 Agilent E4405B Performance Verification Test Record
Agilent Technologies
Model E4405B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty