Agilent Technologies E1401-90493 Work Light User Manual


 
608 Chapter 3
Performance Verification Test Records
Agilent E4408B Performance Verification Test Record
Resolution Bandwidth
3 kHz –0.3 dB (1)________ 0.3 dB
±0.048 dB
9 kHz –0.3 dB (2)________ 0.3 dB
±0.048 dB
10 kHz –0.3 dB (3)________ 0.3 dB ±0.048 dB
30 kHz –0.3 dB (4)________ 0.3 dB
±0.048 dB
100 kHz –0.3 dB (5)________ 0.3 dB
±0.048 dB
120 kHz –0.3 dB (6)________ 0.3 dB
±0.048 dB
300 kHz –0.3 dB (7)________ 0.3 dB
±0.048 dB
1 MHz –0.3 dB (8)________ 0.3 dB
±0.048 dB
3 MHz –0.3 dB (9)________ 0.3 dB
±0.048 dB
5 MHz –0.6 dB (10)________ 0.6 dB
±0.076 dB
300 Hz (Option 1DR only)
3.0 dB (11)________ 0.3 dB ±0.048 dB
200 Hz (Option 1DR only)
3.0 dB (12)________ 0.3 dB ±0.048 dB
100 Hz (Option 1DR only)
3.0 dB (13)________ 0.3 dB ±0.048 dB
18. Absolute Amplitude
Accuracy (Reference
Settings)
Log, Preamp Off –0.4 dB (1)________ 0.4 dB ±0.09 dB
Lin, Preamp Off –0.4 dB (2)________ 0.4 dB
±0.09 dB
20. Overall Absolute
Amplitude Accuracy
0 dBm Reference Level
0 dBm input –0.6 dB (1)________ 0.6 dB
±0.234 dB
–10 dBm input –0.6 dB (2)________ 0.6 dB
±0.115 dB
–20 dBm input –0.6 dB (3)________ 0.6 dB ±0.110 dB
–30 dBm input –0.6 dB (4)________ 0.6 dB
±0.109 dB
Table 3-14 Agilent E4408B Performance Verification Test Record
Agilent Technologies
Model E4408B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty