Chapter 3 545
Performance Verification Test Records
Agilent E4405B Performance Verification Test Record
ac Coupled
Band 0A
(800 MHz to 1.0 GHz)
(serial number
US39440327 or greater)
Maximum Response (13)________ 1.0 dB
±0.11 dB
Minimum Response –1.0 dB (14)________
±0.11 dB
Peak-to-Peak Response (15)________ 2.0 dB ±0.11 dB
ac Coupled
Band 0B
(1.7 GHz to 2.0 GHz)
(serial number
US39440327 or greater)
Maximum Response (16)________ 1.0 dB
±0.11 dB
Minimum Response –1.0 dB (17)________
±0.11 dB
Peak-to-Peak Response (18)________ 2.0 dB
±0.11 dB
ac Coupled
Band 1
(3.0 GHz to 6.7 GHz)
Maximum Response (19)________ 2.5 dB
±0.39 dB
Minimum Response –2.5 dB (20)________ ±0.39 dB
Peak-to-Peak Response (21)________ 3.0 dB
±0.39 dB
ac Coupled
Band 2
(6.7 GHz to 13.2 GHz)
Maximum Response (22)________ 3.0 dB
±0.40 dB
Minimum Response –3.0 dB (23)________
±0.40 dB
Table 3-10 Agilent E4405B Performance Verification Test Record
Agilent Technologies
Model E4405B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty