Agilent Technologies E1401-90493 Work Light User Manual


 
382 Chapter 2
Performance Verification Tests
50. Gate Delay Accuracy and Gate Length Accuracy: Agilent E4401B, E4402B, E4404B,
E4405B, and E4407B (Option 1D6)
2-106.
Record the
Time (1)-(2) max value as the Maximum Gate Delay in
Table 2-106.
13.Clear all current oscilloscope measurements. Click on
Measure, Clear,
and
All Measurements.
14.Adjust the oscilloscope horizontal position to center the pulse on
Channel 2 on the display. Switching the trigger source to channel 2
will help with the measurement.
15.Activate the pulse width measurement function for channel 2 of the
oscilloscope.
Click on
Measure, Time, and click on +width.
Set the Source to be channel 2 and click
OK.
16.Record the +width (2) mean value in Table 2-106 as the 1
µs Gate
Length.
17.On the analyzer, press
Sweep, Sweep Time
150 ms
, Sweep
Gate Setup
(or Gate), Edge Setup (or Edge Gate)
Gate Delay, 10ms
Sweep
, Gate Setup
Edge Setup
(or Edge Gate), Gate Length, 65 ms
18.Set the universal counter controls as follows:
19.Record the universal counter readout value as the 65ms Gate Length
in Table 2-106.
Function Pos Width 1
Gate Time 0.1 s
Auto Trigger On
Channel 1
Coupling DC
Impedance 1 M
X10 Atten Off
100 kHz Filter Off
Common 1 On