Agilent Technologies E1401-90493 Work Light User Manual


 
Chapter 2 357
Performance Verification Tests
45. Tracking Generator Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option
1DN or 1DQ)
Figure 2-65 Harmonic Spurious Outputs Test Setup
2. Complete this step only if more than 24 hours have elapsed since
performing a front-panel calibration of the microwave analyzer.
The microwave analyzer should be allowed to warm up for at least 5
minutes before proceeding.
Complete a front-panel calibration of the microwave analyzer by
performing the following steps:
a. Preset the microwave analyzer.
b. Connect a BNC cable between CAL OUTPUT and 50
Input.
c. Press
CAL, REALIGN LO & IF.
d. Set
FREQUENCY, 300 MHz.
e. Set
SPAN, 20 MHz.
f. Set
AMPLITUDE, –10 dBm.
g. Press
PEAK SEARCH.
h. Press
CAL, REF LVL ADJ and use the ⇑ ⇓ arrows to adjust the DAC
value to a marker amplitude reading of
10 dBm. Press
STORE REF LVL.
i. Disconnect the BNC cable from between the CAL OUTPUT and
50
Input.
3. Press
Preset on the analyzer under test. Press the Factory Preset
softkey, if it is displayed.
4. Set the analyzer by pressing the following keys:
FREQUENCY, 10 MHz
SPAN
, Zero Span
BW/Avg
, 10 kHz
Source
, Amplitude (On)