Agilent Technologies E1401-90493 Work Light User Manual


 
Chapter 2 293
Performance Verification Tests
33. Gain Compression: Agilent E4401B, E4402B, E4403B, and E4411B
10.Set the synthesized sweeper amplitude Off.
11.Set the synthesized signal generator amplitude to
24 dBm.
12.Press
Preset on the analyzer. Press the Factory Preset softkey, if it is
displayed. Press
System, Alignments, Auto Align, Off. Set the analyzer
by pressing the following keys:
FREQUENCY, Center Freq, 50 MHz (or as indicated in Table 2-78)
SPAN, 150 kHz (or as indicated in Table 2-78)
AMPLITUDE, More, Y Axis Units (or Amptd Units), dBm
AMPLITUDE
, Ref Level, –10 dBm, Attenuation 0 dB
AMPLITUDE
, Scale/Div, 10 dB
BW/Avg
, Res BW, 30 kHz (or as indicated in Table 2-78)
BW/Avg, Video BW, 300 Hz (or as indicated in Table 2-78)
13.On the synthesized sweeper, set the appropriate power level to the
setting recorded in Table 2-77. Then set RF to Off.
75
Input only: Adjust the power level for a 2.0 dBm reading.
14.On the analyzer, press
Peak Search (or Search).
15.Adjust the amplitude of the synthesized signal generator to achieve
a marker amplitude reading within 0.5 dB of the value indicated in
Table 2-77. The marker amplitude is the uncompressed amplitude.
16.On the analyzer, press the following keys:
Peak Search (or Search)
Marker
Delta
17.On the synthesized sweeper, set RF to On. The amplitude should be
the same as recorded in Table 2-77.
18.On the analyzer, press
Peak Search (or Search). This is the
compressed amplitude. The marker delta (
Mkr1) amplitude is the
measured gain compression.
19.Record the measured gain compression in the performance test
record as the Entry listed in Table 2-78.
20.Repeat step 6 through step 19 for each set of settings in Table 2-77
and Table 2-78.