Performance Verification Test Records
Agilent E4403B Performance Verification Test Record
23. Frequency Response Note: Enter data in the appropriate section below depending upon
the ambient temperature at which the test was performed.
20 to 30
° C
Maximum Response (1)________ 0.5 dB
±0.19 dB
Minimum Response –0.5 dB (2)________ ±0.19 dB
Peak-to-Peak Response (3)________ 1.0 dB
±0.19 dB
0 to 55
° C
Maximum Response (1)________ 1.0 dB
±0.19 dB
Minimum Response –1.0 dB (2)________
±0.19 dB
Peak-to-Peak Response (3)________ 2.0 dB
±0.19 dB
29. Other Input-Related
Spurious Responses
Input Frequency
2042.8 MHz (1)________ –65 dBc
±0.891 dB
2642.8 MHz (2)________ –65 dBc
±0.891 dB
1820.8 MHz (3)________ –65 dBc
±0.891 dB
278.5 MHz (4)________ –65 dBc
±0.891 dB
31. Spurious Responses Note: Entry 2 does not apply to the Agilent E4403B.
300 MHz TOI 7.5 dBm (1)________
±0.93 dB
300 MHz SHI 30 dBm (3)________
±0.41 dB
900 MHz SHI 40 dBm (4)________ ±0.41 dB
33. Gain Compression Note: Entry 2 does not apply to the Agilent E4403B.
Test Frequency
53 MHz (1)________ 1.0 dB
±0.123 dB
1403 MHz (3)________ 1.0 dB
±0.123 dB
2503 MHz (4)________ 1.0 dB
±0.123 dB
Table 3-6 Agilent E4403B Performance Verification Test Record
Agilent Technologies
Model E4403B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty