552 Chapter 3
Performance Verification Test Records
Agilent E4405B Performance Verification Test Record
Negative Level Flatness,
1 MHz to 10 MHz
–3.0 dB (4)________ ±0.142 dB
Positive Level Flatness,
>10 MHz to 1.5 GHz
(5)________ 2.0 dB
±0.142 dB
Negative Level Flatness,
>10 MHz to 1.5 GHz
–2.0 dB (6)________
±0.142 dB
Positive Level Flatness,
>1.5 GHz
(7)________ 2.0 dB
±0.142 dB
Negative Level Flatness,
>1.5 GHz
–2.0 dB (8)________
±0.142 dB
46. Tracking Generator
Harmonic Spurious
Outputs
(Option 1DN only)
2
nd
Harmonic, <20 kHz
(1)________ –15 dBc
±2.6 dB
2
nd
Harmonic, 20 kHz
(2)________ –25 dBc
±2.6 dB
3
rd
Harmonic, <20 kHz
(3)________ –15 dBc
±2.6 dB
3
rd
Harmonic, 20 kHz
(4)________ –25 dBc
±2.6 dB
48. Tracking Generator
Non-Harmonic
Spurious Outputs
(Option 1DN only)
Highest Non-Harmonic
Spurious Output
Amplitude, 9 kHz to 2 GHz
(1)________ –27 dBc
±2.63 dB
Highest Non-Harmonic
Spurious Output
Amplitude, 2 GHz to
3GHz
(2)________ –23 dBc
±3.14 dB
49. Tracking Generator
L.O. Feedthrough
(Option 1DN only)
Table 3-10 Agilent E4405B Performance Verification Test Record
Agilent Technologies
Model E4405B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty