Agilent Technologies E1401-90493 Work Light User Manual


 
500 Chapter 3
Performance Verification Test Records
Agilent E4404B Performance Verification Test Record
1 ms (Option AYX or B7D
only)
–1.0% (6)________ 1.0% ±0.16%
500
µs (Option AYX or
B7D only)
–1.0% (7)________ 1.0% ±0.16%
100
µs (Option AYX or
B7D only)
–1.0% (8)________ 1.0% ±0.16%
12. Display Scale Fidelity
Cumulative Log Fidelity,
Res BW
1 kHz
dB from Ref Level
–4 –0.30 dB (1)________ 0.30 dB
±0.082 dB
–8 –0.30 dB (2)________ 0.30 dB
±0.078 dB
–12 –0.40 dB (3)________ 0.40 dB
±0.075 dB
–16 –0.40 dB (4)________ 0.40 dB
±0.073 dB
–20 –0.40 dB (5)________ 0.40 dB
±0.078 dB
–24 –0.50 dB (6)________ 0.50 dB
±0.074 dB
–28 –0.50 dB (7)________ 0.50 dB
±0.073 dB
–32 –0.60 dB (8)________ 0.60 dB ±0.077 dB
–36 –0.60 dB (9)________ 0.60 dB
±0.075 dB
–40 –0.60 dB (10)________ 0.60 dB
±0.081 dB
–44 –0.70 dB (11)________ 0.70 dB ±0.077 dB
–48 –0.70 dB (12)________ 0.70 dB
±0.076 dB
–52 –0.70dB (13)________ 0.70 dB
±0.080 dB
–56 –0.70 dB (14)________ 0.70 dB ±0.078 dB
–60 –0.70 dB (15)________ 0.70 dB
±0.084 dB
–64 –0.80 dB (16)________ 0.80 dB
±0.081 dB
–68 –0.80 dB (17)________ 0.80 dB ±0.080 dB
Table 3-8 Agilent E4404B Performance Verification Test Record
Agilent Technologies
Model E4404B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty