Performance Verification Test Records
Agilent E4403B Performance Verification Test Record
6. Frequency Span
Accuracy
Span Start Freq
3000 MHz 0 Hz 2370 MHz (1)________ 2430 MHz
±6.12 MHz
100 MHz 10 MHz 79 MHz (2)________ 81 MHz ±204 kHz
100 kHz 10 MHz 79 kHz (3)________ 81 kHz
±204 Hz
100 MHz 800 MHz 79 MHz (4)________ 81 MHz
±204 kHz
100 kHz 800 MHz 79 kHz (5)________ 81 kHz
±204 Hz
100 MHz 1400 MHz 79 MHz (6)________ 81 MHz
±204 kHz
100 kHz 1499 MHz 79 kHz (7)________ 81 kHz
±204 Hz
7. Noise Sidebands
Offset from 1 GHz signal
10 kHz (1)________ –90 dBc/Hz
±2.44 dB
20 kHz (2)________ –100 dBc/Hz
±2.44 dB
30 kHz (3)________ –106 dBc/Hz
±2.44 dB
9. System-Related
Sidebands
Offset from 500 MHz
signal
30 kHz to 230 kHz (1)________ –65 dBc
±1.30 dB
–30 kHz to –230 kHz (2)________ –65 dBc
±1.30 dB
10. Residual FM
1 kHz Res BW (1)________ 150 Hz
±21 Hz
11. Sweep Time Accuracy
Sweep Time
5 ms –1.0% (1)________ 1.0%
±0.16%
Table 3-6 Agilent E4403B Performance Verification Test Record
Agilent Technologies
Model E4403B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty