Agilent Technologies E1401-90493 Work Light User Manual


 
Chapter 3 629
Performance Verification Test Records
Agilent E4411B Performance Verification Test Record
9 kHz 7.2 kHz (11)________ 10.8 kHz ±163 Hz
22. Frequency Response Note: Enter data in the appropriate section below depending upon
the ambient temperature at which the test was performed and
installed options.
20 to 30
° C
50
Input (No Option
1DP)
(9 kHz to 1.5 GHz)
Maximum Response (1)________ 0.50 dB
±0.19 dB
Minimum Response –0.50 dB (2)________
±0.19 dB
Peak-to-Peak Response (3)________ 1.0 dB
±0.19 dB
Option 1DP (75
)
(1 MHz to 1.5 GHz)
Maximum Response (1)________ 0.50 dB
±0.11 dB
Minimum Response –0.50 dB (2)________
±0.11 dB
Peak-to-Peak Response (3)________ 1.0 dB
±0.11 dB
0 to 55
° C
50
Input (No Option
1DP)
(9 kHz to 1.5 GHz)
Maximum Response (1)________ 1.0 dB
±0.19 dB
Minimum Response –1.0 dB (2)________
±0.19 dB
Peak-to-Peak Response (3)________ 2.0 dB ±0.19 dB
Option 1DP (75
)
(1 MHz to 1.5 GHz)
Maximum Response (1)________ 1.0 dB
±0.11 dB
Table 3-16 Agilent E4411B Performance Verification Test Record
Agilent Technologies
Model E4411B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty