444 Chapter 3
Performance Verification Test Records
Agilent E4401B Performance Verification Test Record
Peak-to-Peak Response (3)________ 4.0 dB ±0.28 dB
28. Other Input-Related
Spurious Responses
Input Frequency
542.8 MHz (1)________ –65 dBc
±0.889 dB
510.7 MHz (2)________ –65 dBc
±0.889 dB
1310.7 MHz (3)________ –45 dBc
±0.889 dB
30. Spurious Responses Note: Enter data in the appropriate section below depending upon the
input impedance and serial number of the analyzer.
50 MHz TOI, 1 kHz RBW,
50
Ω (Serial Number less
than US39440413)
10 dBm (1)________
±0.93 dB
50 MHz TOI, 1 kHz RBW,
50
Ω (Serial Number
US39440413 or greater)
13.5 dBm (1)________
±0.93 dB
50 MHz TOI, 1 kHz RBW,
75
58.75 dBmV (1)________ ±0.93 dB
50 MHz TOI, 30 Hz RBW,
50
(Option 1DR only)
(
Serial Number less than
US39440413)
10 dBm (2)________
±0.93 dB
50 MHz TOI, 30 Hz RBW,
50
(Option 1DR only)
(
Serial Number
US39440413 or greater)
13.5 dBm (2)________
±0.93 dB
50 MHz TOI, 30 Hz RBW,
75
(Option 1DR only)
58.75 dBmV (2)________
±0.93 dB
40 MHz SHI, 50
35 dBm (3)________ ±0.50 dB
40 MHz SHI, 75
83.75 dBmV (3)________ ±0.50 dB
33. Gain Compression
Test Frequency
Table 3-2 Agilent E4401B Performance Verification Test Record
Agilent Technologies
Model E4401B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty