Chapter 3 567
Performance Verification Test Records
Agilent E4407B Performance Verification Test Record
Incremental Log Fidelity,
Res BW
300 Hz
(Option 1DR only)
dB from Ref Level
–4 –0.4 dB (68)________ 0.4 dB
±0.082 dB
–8 –0.4 dB (69)________ 0.4 dB
±0.078 dB
–12 –0.4 dB (70)________ 0.4 dB ±0.075 dB
–16 –0.4 dB (71)________ 0.4 dB
±0.073 dB
–20 –0.4 dB (72)________ 0.4 dB
±0.078 dB
–24 –0.4 dB (73)________ 0.4 dB
±0.074 dB
–28 –0.4 dB (74)________ 0.4 dB
±0.073 dB
–32 –0.4 dB (75)________ 0.4 dB
±0.077 dB
–36 –0.4 dB (76)________ 0.4 dB
±0.075 dB
–40 –0.4 dB (77)________ 0.4 dB
±0.081 dB
–44 –0.4 dB (78)________ 0.4 dB
±0.077 dB
–48 –0.4 dB (79)________ 0.4 dB
±0.076 dB
–52 –0.4 dB (80)________ 0.4 dB
±0.080 dB
–56 –0.4 dB (81)________ 0.4 dB
±0.078 dB
–60 –0.4 dB (82)________ 0.4 dB ±0.084 dB
–64 –0.4 dB (83)________ 0.4 dB
±0.081 dB
–68 –0.4 dB (84)________ 0.4 dB
±0.080 dB
–72 –0.4 dB (85)________ 0.4 dB ±0.084 dB
–76 –0.4 dB (86)________ 0.4 dB
±0.083 dB
–80 –0.4 dB (87)________ 0.4 dB
±0.089 dB
Linear Fidelity,
Res BW
1 kHz
dB from Ref Level
Table 3-12 Agilent E4407B Performance Verification Test Record
Agilent Technologies
Model E4407B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty