Agilent Technologies E1401-90493 Work Light User Manual


 
Chapter 3 555
Performance Verification Test Records
Agilent E4405B Performance Verification Test Record
20 dBm –0.77 dB (20)________ 0.77 dB ±0.161 dB
−30 dBm –0.70 dB (21)________ 0.70 dB ±0.156 dB
−40 dBm –0.75 dB (22)________ 0.75 dB ±0.161 dB
50 dBm –0.91 dB (23)________ 0.91 dB ±0.152 dB
−60 dBm –1.05 dB (24)________ 1.05 dB ±0.161 dB
cdmaOne Receive
Channel Power, Preamp
Off
(Option BAC only)
Cellular Band
Input Amplitude
40 dBm –1.46 dB (25)________ 1.46 dB ±0.161 dB
60 dBm –1.46 dB (26)________ 1.46 dB ±0.161 dB
70 dBm –1.46 dB (27)________ 1.46 dB ±0.161 dB
80 dBm –1.46 dB (28)________ 1.46 dB ±0.161 dB
85 dBm –1.46 dB (29)________ 1.46 dB ±0.152 dB
PCS Band
Input Amplitude
40 dBm –1.35 dB (30)________ 1.35 dB ±0.161 dB
60 dBm –1.35 dB (31)________ 1.35 dB ±0.161 dB
−70 dBm –1.35 dB (32)________ 1.35 dB ±0.161 dB
−80 dBm –1.35 dB (33)________ 1.35 dB ±0.161 dB
85 dBm –1.35 dB (34)________ 1.35 dB ±0.152 dB
cdmaOne Receive
Channel Power, Preamp
On
(Option BAC only)
Cellular Band
Input Amplitude
Table 3-10 Agilent E4405B Performance Verification Test Record
Agilent Technologies
Model E4405B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty