Agilent Technologies E1401-90493 Work Light User Manual


 
632 Chapter 3
Performance Verification Test Records
Agilent E4411B Performance Verification Test Record
Negative Vernier
Accuracy
–0.75 dB (3)________ ±0.008 dB
Power Sweep Accuracy (4)________ 1.5 dB
±0.008 dB
75
(Option 1DQ)
Absolute Amplitude
Accuracy
–1.5 dB (1)________ 1.5 dB ±0.096 dB
Positive Vernier Accuracy (2)________ 0.9 dB
±0.008 dB
Negative Vernier
Accuracy
–0.9 dB (3)________
±0.008 dB
Power Sweep Accuracy (4)________ 1.8 dB
±0.008 dB
43. Tracking Generator
Level Flatness
Note: Enter data in the appropriate section below depending upon
the input impedance and serial number of the analyzer.
50
(Option 1DN)
Positive Level Flatness,
<1 MHz
(1)________ 2.0 dB
±0.73 dB
Negative Level Flatness,
<1 MHz
–2.0 dB (2)________
±0.73 dB
Positive Level Flatness,
1 MHz to 10 MHz
(3)________ 2.0 dB
±0.19 dB
Negative Level Flatness,
1 MHz to 10 MHz
–2.0 dB (4)________
±0.19 dB
Positive Level Flatness,
>10 MHz
(5)________ 1.5 dB
±0.19 dB
Negative Level Flatness,
>10 MHz
–1.5 dB (6)________
±0.19 dB
75
(Option 1DQ)
Positive Level Flatness,
1 MHz to 10 MHz
(3)________ 2.5 dB
±0.19 dB
Negative Level Flatness,
1 MHz to 10 MHz
–2.5 dB (4)________
±0.19 dB
Table 3-16 Agilent E4411B Performance Verification Test Record
Agilent Technologies
Model E4411B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty