Agilent Technologies E1401-90493 Work Light User Manual


 
448 Chapter 3
Performance Verification Test Records
Agilent E4401B Performance Verification Test Record
Power Sweep Accuracy (4)________ 1.8 dB ±0.008 dB
43. Tracking Generator
Level Flatness
Note: Enter data in the appropriate section below depending upon the
input impedance of the analyzer.
(Option 1DN or Option
1DQ only)
Positive Level Flatness,
<1 MHz
(1)________ 2.0 dB
±0.73 dB
Negative Level Flatness,
<1 MHz
–2.0 dB (2)________
±0.73 dB
Positive Level Flatness,
1 MHz to 10 MHz
(3)________ 2.0 dB
±0.190 dB
Negative Level Flatness,
1 MHz to 10 MHz
–2.0 dB (4)________
±0.190 dB
Positive Level Flatness,
>10 MHz
(5)________ 1.5 dB
±0.190 dB
Negative Level Flatness,
>10 MHz
–1.5 dB (6)________
±0.190 dB
75
(Option 1DQ)
Positive Level Flatness,
1 MHz to 10 MHz
(3)________ 2.5 dB
±0.190 dB
Negative Level Flatness,
1 MHz to 10 MHz
–2.5 dB (4)________
±0.190 dB
Positive Level Flatness,
>10 MHz
(5)________ 2.0 dB
±0.190 dB
Negative Level Flatness,
>10 MHz
–2.0 dB (6)________
±0.190 dB
45. Tracking Generator
Harmonic Spurious
Outputs (Option 1DN or
Option 1DQ only)
2
nd
Harmonic, <20 MHz
(1)________ –20 dBc
±2.6 dB
Table 3-2 Agilent E4401B Performance Verification Test Record
Agilent Technologies
Model E4401B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty