Agilent Technologies E1401-90493 Work Light User Manual


 
354 Chapter 2
Performance Verification Tests
44. Tracking Generator Level Flatness: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,
and E4408B (Option 1DN)
2. Set the analyzer resolution bandwidth to 10 kHz by pressing
BW/Avg, Res BW, 10 kHz.
Repeat step 3 through step 7 for each Center Frequency value in
Table 2-98.
3. Set the analyzer center frequency to 9 kHz, by pressing
FREQUENCY,
9 kHz.
4. Press
Single.
5. Record the DVM readout in Table 2-98.
6. Subtract the 100 kHz Level Flatness readout in Table 2-97 from the
100 kHz DVM Readout in Table 2-98 and record as the DVM Offset
at 100 kHz.
DVM Offset at 100 kHz ___________ dB
For example, if the Level Flatness reading from Table 2-97 is
0.7 dB and the DVM Readout from Table 2-98 is
0.53 dBm, the
DVM offset would be
1.23 dB.
7. Add the DVM Offset at 100 kHz from step 6, above, to each of the
DVM Readouts in Table 2-98 and record as the Corrected Level
Flatness in Column 3.
For example, if the DVM Readout from Table 2-98 is 0.22 dBm,
and the DVM Offset is
1.23 dB, the Corrected Level Flatness
would be
1.01 dB.
8. Press
System, Alignments, Auto Align, All.
Set to 50
impedance
SMATH 10
a
Set to dBm
MATH 5
a
Set to Synchronous
SETACV 3
a
Sub-sampled mode
a. To set the Agilent 3458A multimeter functions from the
front panel, press the blue shift key, then Recall State (T)
key. Use the
(up) and (down) arrows to select the
appropriate function, then enter the value from the
numeric keypad and press enter.
Parameter Setting
DVM Offset DVM Readout= Level Flatness
Corrected Level Flatness DVM Readout= DVM Offset+