Agilent Technologies E1401-90493 Work Light User Manual


 
574 Chapter 3
Performance Verification Test Records
Agilent E4407B Performance Verification Test Record
1 kHz 850 Hz (9)________ 1.15 kHz ±22 Hz
120 kHz 96 kHz (10)________ 144 kHz
±2.17 kHz
9 kHz 7.2 kHz (11)________ 10.8 kHz
±163 Hz
24. Frequency Response Note: Enter data in the appropriate section below depending upon the
ambient temperature at which the test was performed and installed
options.
20 to 30
° C
Non-Option UKB
dc Coupled
Band 0
(9 kHz to 3.0 GHz)
Maximum Response (1)________ 0.46 dB
±0.19 dB
Minimum Response –0.46 dB (2)________
±0.19 dB
Peak-to-Peak Response (3)________ 0.92 dB
±0.19 dB
Option UKB
dc Coupled
Band 0
(100 Hz to 3.0 GHz
Maximum Response (1)________ 0.5 dB
±0.19 dB
Minimum Response –0.5 dB (2)________
±0.19 dB
Peak-to-Peak Response (3)________ 1.0 dB ±0.19 dB
Non-Option UKB
dc Coupled
Band 0A
(800 MHz to 1.0 GHz)
(serial number
US39440871 or greater)
Maximum Response (4)________ 0.46 dB
±0.11 dB
Minimum Response –0.46 dB (5)________
±0.11 dB
Table 3-12 Agilent E4407B Performance Verification Test Record
Agilent Technologies
Model E4407B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty