Chapter 3 475
Performance Verification Test Records
Agilent E4402B Performance Verification Test Record
50. Gate Delay Accuracy
and Gate Length
Accuracy
(Option 1D6 only)
Minimum Gate Delay 499.9 ns (1)________ 1.5001
µs ±549 ps
Maximum Gate Delay 499.9 ns (2)________ 1.5001
µs ±549 ps
1 µs Gate Length 499.9 ns (3)________ 1.5001 µs ±520 ps
65 ms Gate Length 64.993 ms (4)________ 65.007 ms
±647 ns
51. Gate Mode Additional
Amplitude Error (Option
1D6 only)
Amplitude Error –0.2 dB (1)________ 0.2 dB
±0.023 dB
58. GSM Phase and
Frequency Error
(Option BAH and B7E)
Peak Phase Error –2.1 Deg (1)________ 2.1 Deg
±0.52 Deg
RMS Phase Error –1.1 Deg (2)________ 1.1 Deg
±0.000063 Deg
Frequency Error –10 Hz (3)________ 10 Hz
±0.006 Hz
59. Comms Absolute Power
Accuracy
(Options BAC or BAH)
20 to 30
° C
cdmaOne Channel Power
Accuracy
(Option BAC only)
Cellular Band
Input Amplitude
15 dBm –0.90 dB (1)________ 0.90 dB
±0.173 dB
5 dBm –0.90 dB (2)________ 0.90 dB ±0.161 dB
25 dBm –0.86 dB (3)________ 0.86 dB ±0.161 dB
Table 3-4 Agilent E4402B Performance Verification Test Record
Agilent Technologies
Model E4402B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty