Agilent Technologies E1401-90493 Work Light User Manual


 
460 Chapter 3
Performance Verification Test Records
Agilent E4402B Performance Verification Test Record
13. Input Attenuation
Switching Uncertainty
Input Attenuation Setting
0 dB –0.3 dB (1)________ 0.3 dB
±0.101 dB
5 dB –0.3 dB (2)________ 0.3 dB ±0.104 dB
15 dB –0.3 dB (3)________ 0.3 dB
±0.102 dB
20 dB –0.3 dB (4)________ 0.3 dB
±0.098 dB
25 dB –0.35 dB (5)________ 0.35 dB
±0.098 dB
30 dB –0.40 dB (6)________ 0.40 dB
±0.096 dB
35 dB –0.45 dB (7)________ 0.45 dB
±0.099 dB
40 dB –0.50 dB (8)________ 0.50 dB
±0.096 dB
45 dB –0.55 dB (9)________ 0.55 dB
±0.099 dB
50 dB –0.60 dB (10)________ 0.60 dB
±0.095 dB
55 dB –0.65 dB (11)________ 0.65 dB
±0.099 dB
60 dB –0.70 dB (12)________ 0.70 dB
±0.100 dB
65 dB –0.75 dB (13)________ 0.75 dB
±0.111 dB
15. Reference Level
Accuracy
Log, Res BW
1 kHz
Reference Level
–10 dBm –0.3 dB (1)________ 0.3 dB
±0.091 dB
0 dBm –0.3 dB (2)________ 0.3 dB ±0.100 dB
–30 dBm –0.3 dB (3)________ 0.3 dB
±0.092 dB
–40 dBm –0.3 dB (4)________ 0.3 dB
±0.094 dB
–50 dBm –0.5 dB (5)________ 0.5 dB ±0.095 dB
–60 dBm –0.5 dB (6)________ 0.5 dB
±0.097 dB
Table 3-4 Agilent E4402B Performance Verification Test Record
Agilent Technologies
Model E4402B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty