Agilent Technologies E1401-90493 Work Light User Manual


 
Chapter 3 593
Performance Verification Test Records
Agilent E4407B Performance Verification Test Record
58. GSM Phase and
Frequency Error
(Option BAH and B7E)
Peak Phase Error –2.1 Deg (1)________ 2.1 Deg 0.52 Deg
RMS Phase Error –1.1 Deg (2)________ 1.1 Deg 0.000063 Deg
Frequency Error –10 Hz (3)________ 10 Hz
±0.006 Hz
59. Comms Absolute Power
Accuracy
(Options BAC or BAH)
20 to 30
° C
cdmaOne Channel Power
Accuracy
(Option BAC only)
Cellular Band
Input Amplitude
15 dBm –0.95 dB (1)________ 0.95 dB
±0.173 dB
5 dBm –0.95 dB (2)________ 0.95 dB ±0.161 dB
25 dBm –0.84 dB (3)________ 0.84 dB ±0.161 dB
45 dBm –0.87 dB (4)________ 0.87 dB ±0.152 dB
55 dBm –0.95 dB (5)________ 0.95 dB ±0.152 dB
70 dBm –1.07 dB (6)________ 1.07 dB ±0.161 dB
PCS Band
Input Amplitude
15 dBm –0.93 dB (7)________ 0.93 dB
±0.173 dB
5 dBm –0.93 dB (8)________ 0.93 dB ±0.161 dB
−25 dBm –0.78 dB (9)________ 0.78 dB ±0.161 dB
45 dBm –0.77 dB (10)________ 0.77 dB ±0.152 dB
55 dBm –0.85 dB (11)________ 0.85 dB ±0.152 dB
70 dBm –0.97 dB (12)________ 0.97 dB ±0.161 dB
Table 3-12 Agilent E4407B Performance Verification Test Record
Agilent Technologies
Model E4407B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty