Agilent Technologies E1401-90493 Work Light User Manual


 
610 Chapter 3
Performance Verification Test Records
Agilent E4408B Performance Verification Test Record
24. Frequency Response Note: Enter data in the appropriate section below depending upon the
ambient temperature at which the test was performed and installed
options.
20 to 30
° C
Band 0
(9 kHz to 3.0 GHz)
Maximum Response (1)________ 0.50 dB
±0.19 dB
Minimum Response –0.50 dB (2)________
±0.19 dB
Peak-to-Peak Response (3)________ 1.0 dB
±0.19 dB
Band 1
(3.0 GHz to 6.7 GHz)
Maximum Response (10)________ 1.5 dB
±0.39 dB
Minimum Response –1.5 dB (11)________
±0.39 dB
Peak-to-Peak Response (12)________ 2.6 dB
±0.39 dB
Band 2
(6.7 GHz to 13.2 GHz)
Maximum Response (13)________ 2.0 dB
±0.40 dB
Minimum Response –2.0 dB (14)________
±0.40 dB
Peak-to-Peak Response (15)________ 3.6 dB
±0.40 dB
Band 3
(13.2 GHz to 26.5 GHz)
Maximum Response (16)________ 2.0 dB
±0.59 dB
Minimum Response –2.0 dB (17)________ ±0.59 dB
Peak-to-Peak Response (18)________ 3.6 dB
±0.59 dB
Table 3-14 Agilent E4408B Performance Verification Test Record
Agilent Technologies
Model E4408B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty