Performance Verification
4-4
AWG710 Service Manual
Figure 4-1: Diagnostic menu
Push the Execute Diagnostic side button to execute all of the AWG710
Arbitrary Waveform Generator diagnostics automatically.
NOTE. Push the Abort Diagnostic button (side menu) to stop a selected diagnos-
tic after it is completed. This button does not stop the diagnostic while it is
processing. The Abort button stops any further diagnostics from operating.
The internal diagnostics do an extensive verification of AWG710 Arbitrary
Waveform Generator functions. While this verification progresses, the screen
displays the clock icon. When finished, the resulting status appears on the
screen.
Verify that no failures are found and reported on-screen. If the diagnostics
terminates without error, Pass is displayed instead of the - - -. If a value is
displayed, meaning an error is detected, consult a qualified service technician for
further assistance.
Push any bottom or menu button (other than UTILITY) to exit the diagnostic
screen.