AWG710 Test Record
Photocopy this test record and use to record the performance test results for your
AWG710 Arbitrary Waveform Generator.
AWG710 Test Record
Instrument Serial Number:ą Certificate Number:ą
Temperature:ą RH %:ą
Date of Calibration:ą Technician:ą
AWG710 Performance Test Minimum Incoming Outgoing Maximum
Operating Mode
Check Cont Mode Pass/Fail __________ __________ Pass/Fail
Check Triggered Mode (when the FORCE
TRIGGER button is pushed)
Pass/Fail __________ __________ Pass/Fail
Check Triggered Mode (with external triggering) Pass/Fail __________ __________ Pass/Fail
Check Gated Mode (when the FORCE TRIGGER
button is pushed) Pass/Fail __________ __________ Pass/Fail
Check Gated Mode (with the gate signal when
the AWG trigger polarity is set to positive) Pass/Fail __________ __________ Pass/Fail
Check Gated Mode (with the gate signal when the
AWG trigger polarity is set to negative) Pass/Fail __________ __________ Pass/Fail
Amplitude and Offset Accuracy (Normal Out), (except option 02)
CH1 Amplitude 20 mV
200 mV
2V
17.60 mV
194.0 mV
1.9580 V
__________
__________
__________
__________
__________
__________
22.40 mV
206.0 mV
2.0420 V
CH1 Amplitude 20 mV
200 mV
2V
17.60 mV
194.0 mV
1.9580 V
__________
__________
__________
__________
__________
__________
22.40 mV
206.0 mV
2.0420 V
CH1 Offset 0 V
+0.5 V
-0.5 V
- 10.0 mV
+ 0.4825 V
- 0.4825 V
__________
__________
__________
__________
__________
__________
+ 10.0 mV
+ 0.5175 V
- 0.5175 V
CH1 Offset 0 V
+0.5 V
-0.5 V
- 10.0 mV
+ 0.4825 V
- 0.4825 V
__________
__________
__________
__________
__________
__________
+ 10.0 mV
+ 0.5175 V
- 0.5175 V