Check that you have the proper electrical connections. The AWG710 Arbitrary
Waveform Generator operates within the following power supply voltage and
frequency ranges:
Line voltage range 100 - 240 V
Line frequency 48-63Hz
Maximum power 220 VA
Connect the proper power cord from the rear panel power connector to the power
system.
NOTE. The AWG710 Arbitrary Waveform Generator is shipped with a 115 V
power cord. If the AWG710 Arbitrary Waveform Generator is to be used with
230 V power, the power cord must be replaced with one appropriate for the
power source used. See Table 2–2 for the available power cord types.
Push the PRINCIPAL POWER SWITCH (shown in Figure 2–1) on the rear
panel of the instrument. Power is now applied to the instrument standby circuitry.
Once the instrument is installed, leave the PRINCIPAL POWER SWITCH on
and use the ON/STBY switch, located on the front panel, to toggle the instru-
ment between ON and STBY.
PRINCIPAL POWER SWITCH
Fuse
Power connector
Figure 2-1: Rear panel power switch, fuse holder, and power connector
Check Voltage Settings