Performance Verification
AWG710 Service Manual
4-31
Amplitude, Offset Accuracy and Rise Time Tests (for option 02)
These procedures check the accuracy of the AWG710 Arbitrary Waveform
Generator option 02 outputs; amplitude and offset.
Equipment
required
A 50ĂW BNCĂcoaxial cable, a 50ĂWĂprecision terminator, an
SMA(Ma)ĆBNC(Fe)Ăadapter, BNC (female)ĆtoĆdual banana adapter, and
a digital multimeter (DMM).
Prerequisites The AWG710 option 02 Arbitrary Waveform Generator must meet the
prerequisites listed on page 4-8.
Do the following steps to install the test hookup and set the test equipment
controls:
1. Use a 50 W BNC coaxial cable, an SMA(Ma)-BNC(Fe) adapter, a 50 W
precision terminator, and a BNC-to-dual banana adapter to connect the
AWG710 Arbitrary Waveform Generator CH1 output to the DMM input
connector (see Figure 4–12).
DMM
50 W BNCĂcoaxial cable
InputĂconnector
+ĂBNCĆtoĆdual banana adapter
+Ă50 W BNCĂcoaxialĂcable
CH1ĂoutputĂconnector
+ĂSMA(Male)ĆBNC(Female) adapter
+ĂBNC 50 WĂprecision terminator
+Ă50 WĂBNC coaxial cable
AWG710 option 02 Arbitrary Waveform Generator
Figure 4-12: Option 02 output amplitude accuracy initial test hookup
2. Set the DMM controls as follows:
Mode . ......................... VDC
Range ......................... 2V
Input . ......................... Front
Check Amplitude
and
DC Offset