Adjustment Procedures
5-10
AWG710 Service Manual
This procedure uses the general purpose knob on the front panel to set the Magic
Frequency parameter.
Equipment
Required
One oscilloscope (Item 1)
One SMA 50 W coaxial cable (Item 4)
SMA (Fe) to BNC (Ma) 50 W adapter (Item 5)
Do the following steps to connect the test equipment and set test equipment
controls:
1. Use a SMA coaxial cable and a SMA (female) to BNC (male) adapter to
connect the waveform generator CH1 Output connector to the CH1 input
connector on the oscilloscope.
SMA 50 W coaxial cable
AWG710 Arbitrary Waveform Generator Oscilloscope (TDS784D)
CH1 input
CH1 output
SMA(Fe)ĆBNC(Ma)
adapter
Figure 5-4: Hookup for the marker level adjustment
2. Set the Oscilloscope controls as follows:
Vertical......................... CH1
CH 1 coupling . ............... DC
CH1scale .................. 200mV/div
CH 1 input impedance . ......... 50W
Horizontal
Sweep . .................... 200ns/div
Trigger
Source . .................... CH1
Coupling . ................... DC
Slope ...................... Positive
Level ...................... 0V
Mode . ..................... Auto
Magic Frequency