Specifications
1-16
AWG710 Service Manual
Certification and Compliances
The certification and compliances for the AWG710 Arbitrary Waveform
Generator are listed in Table 1–22. See Table 1–23 for installation categories and
pollution degree descriptions.
Table 1-22: Certifications and compliances
Characteristics Description
EC declaration of conformity EC council EMC Directive 89/336/EEC, amended by 89/336/EEC; EN61326-1: 1997
Product Family Standard for Electrical Equipment for Measurement, Control, and
Laboratory Use-EMC Requirements.
Emissions:
EN 61326
EN 61000Ć3Ć2
EMC requirements for class A electrical equipment for
measurement control and laboratory use.
AC power line harmonic emissions
Immunity:
IEC 61000Ć4Ć2
IEC 61000Ć4Ć3
IEC 61000Ć4Ć4
6
IEC 61000Ć4Ć5
IEC 61000Ć4Ć6
IEC 61000Ć4Ć11
Electrostatic Discharge Immunity
Radiated RF Electromagnetic Field Immunity
Electrical Fast Transient/Burst Immunity
Surge Immunity
Conducted Disturbances Induced by RF Field Immunity
Power Line Interruption Immunity
Compliance was demonstrated to the following specification as listed in the Official Journal
of the European Communities:
Low Voltage Directive 73/23/EEC, amended by 93/68/EEC
EN 61010Ć1/A2:1995 Safety requirements for electrical equipment for measureĆ
ment, control, and laboratory use
Australia/New Zealand declaration of
conformity - EMC
Conforms with the following standards in accordance with the Electromagnetic CompatibiliĆ
ty Framework:
AS/NZS 2064.1/2 Class A radiated and Conducted Emissions
Safety
ThirdĂpartyĂcertification
SelfĂdeclaration
UL3111Ć1 - Standard for electrical measuring and test equipment
CAN/CSA C22.2 No. 1010.1 - Safety requirements for electrical equipment for measureĆ
ment, control and laboratory use
IEC 61010Ć1/A2:1995 - Safety requirements for electrical equipment for measurement,
control, and laboratory use
Installation category Power input Ć Installation Category II (as defined in IEC 61010Ć1, Annex J)
Pollution degree Pollution Degree 2 (as defined in IEC 61010Ć1)
6
Up to 200 mV
pĆp