Adjustment Procedures
AWG710 Service Manual
5-9
9. Follow the substeps below to adjust the Marker 2 level:
a. Press Marker 2 Adj. Level
b. Turn the general purpose knob to adjust the waveform of marker low
level voltage to 0.00 V within"50 mV on the oscilloscope display.
10. Press Save to save the setting.
11. Disconnect the oscilloscope.