Performance Verification
4-10
AWG710 Service Manual
The following steps explain how to load files from the Performance Check/Ad-
justment disk into the waveform memory and/or sequence memory.
1. Insert the disk into the AWG710 Arbitrary Waveform Generator floppy disk
drive.
2. Select SETUP (front)!Waveform/Sequence (bottom)!Load... (side)!
Drive... (side).
The Select Drive dialog box appears as shown in Figure 4–4.
Figure 4-4: Loading file; selecting storage drive
3. Select Floppy from the dialog box using the general purpose knob, and then
push the OK side button.
The Select Drive dialog box disappears and the files in the floppy disk are
listed on the Select File dialog box.
4. Use the general purpose knob to select a file to load from the dialog box, and
then push the OK side button.
The waveform or sequence you selected is loaded into the instrument, and
the instrument is also setup with the parameters stored in that file.
5. Remove the floppy disk from the floppy disk drive if the floppy disk is no
longer needed.
6. Push any bottom button or menu button to exit the menu.
Loading Files