Chapter 2 Function Reference — LPM16_Calibrate
©
National Instruments Corporation 2-279 NI-DAQ FRM for PC Compatibles
LPM16_Calibrate
Format
status = LPM16_Calibrate (deviceNumber)
Purpose
Calibrates the LPM devices converter. The calibration calculates the correct offset voltage for
the voltage comparator, adjusts positive linearity and full-scale errors to less than ±0.5 LSB
each, and adjusts zero error to less than ±1 LSB.
Parameters
Input
Using This Function
When the function is called, the ADC1241 ADC goes into a self-calibration cycle. The
function does not return until the self-calibration is completed.
Name Type Description
deviceNumber I16 assigned by configuration utility