BANDWIDTH FILTER ADJUSTMENTS (Cont'd)
TERMINAL CONNECTORS
FIGURE
5-6.
CRYSTAL SHORT CONFIGURATION
PROCEDURE:
NOTE
Allow
30
minutes warmup time before performing adjustments.
1.
Set spectrum analyzer controls as follows:
FREQUENCYBAND
..................................................
.Ol-3GHz
TUNING
.................................................................
35MHz
FREQSPAN/DIV
.........................................................
200kHz
RESOLUTIONBW
.........................................................
lMHz
INPUTATTEN
.............................................................
10dB
REFERENCELEVEL
.......................................................
OdBm
Amplitude Scale
..............................................................
LIN
SWEEPTIME/DIV
........................................................
10msec
SWEEPTRIGGER
.....................................................
FREERUN
Crystal Alignment
2.
Connect equipment as shown in Figure 5
-
5.
NOTE
If Sweep Generator A9 has been replaced or adjusted, perform steps
3
through
8.
If not, proceed to step 9.